AN EVALUATION OF THE BD BACTEC™ FX AUTOMATED BLOOD CULTURE SYSTEM IN PRETORIA, SOUTH AFRICA

Vhudzani Tshisevhe, Ramatlhwa Charlotte Kekana, Nontombi Mbelle

Abstract


The high mortality associated with blood stream infections (BSI) calls for improved and timely diagnosis. This will ensure early initiation of appropriate antimicrobial agents as advocated by several international bodies including the surviving sepsis campaign. Blood culture remains the most important diagnostic method for detection of BSI. The study sought to evaluate a new automated blood culture monitoring system from BD (BD BACTEC™ FX). Twenty commonly isolated microorganisms including fungi and anaerobes were inoculated into blood culture bottles containing 10mls of sheep blood. Both the BD BACTEC™ FX and BacT/Alert systems were loaded with their respective blood culture bottles for comparison. The BD BACTECFX system generally shows earlier time to detection of microbial growth than the BacT/ALERT system. Time to positivity was clearly defined with the BD BACTECFX system. The BD BACTEC FX instrument was found to be as user friendly as the current BacT/ALERT system. Although the workflow was more efficient with the BD BACTEC FX instrument. The early time to positivity may improve patient morbidity and mortality as it may facilitate early initiation of appropriate antimicrobial agent. This ability to influence patient care by the BD BACTEC FX means that it may be the better of the two systems. However, a larger study utilizing clinical specimen is required to confirm this.

 


Keywords


Continuous blood culture monitoring instrument, bloodstream infections, evaluation of blood culture instrument, BacT/ALERT, BD BACTEC FX, verification, validation

Full Text:

PDF

References


Engel C, Brunkhorst FM, Bone HG et al (2007) Epidemiology of sepsis in Germany: results from a national prospective multicentre study. Intensive Care Med 33:606–618.

Suetens C (2008) Surveillance of nosocomial bloodstream infections in Belgian hospitals: results from the national surveillance network, 1992–2004 and 2007. Scientific Institute of Public Health, Brussels.

Dellinger RP, Levy MM, Rhodes A et al (2013) Surviving sepsis campaign: international guidelines for management of severe sepsis and septic shock: 2012. Critical Care Medicine 41(2):580-637.

Bass SN, Bauer SR, Neuner EA, Lam SW (2015) Mortality Risk Factors for Critically Ill Patients with Carbapenem-Resistant Bacteremia: Impact of Combination Antimicrobial Therapy. Antimicrobial agents and chemotherapy 59(7):3748-3753.

Rohner P, Auckenthaler R (1999) Review on evaluations of currently available blood‐culture systems. Clinical microbiology and infection 5(9): 513-529.

Flayhart D, Borek AP, Wakefield T, Dick J, Carroll KC (2007) Comparison of BACTEC PLUS blood culture media to BacT/Alert FA blood culture media for detection of bacterial pathogens in samples containing therapeutic levels of antibiotics. Journal of clinical microbiology 45(3): 816-821.

BD diagnostics (2008) BD BACTECTM FX brochure. Heidelberg, Germany.

Clark RB, Lewinski MA, Loeffelholz MJ, Tibbetts RJ (2009) Cumitech 31A, Verification and Validation of Procedures in the Clinical Microbiology Laboratory. Coordinating ed., S. E. Sharp. ASM Press, Washington, DC.

Rosa RG, Goldani LZ (2014) Cohort study of the impact of time to antibiotic administration on mortality in patients with febrile neutropenia. Antimicrobial agents and chemotherapy 58(7):3799-803.

Lodise TP, McKinnon PS, Swiderski L, Rybak MJ (2003) Outcomes analysis of delayed antibiotic treatment for hospital-acquired Staphylococcus aureus bacteremia. Clinical Infectious Diseases 36(11):1418-23.

Hughes JG, Vetter EA, Patel R et al (2001) Culture with BACTEC Peds Plus/F bottles compared with conventional methods for detection of bacteria in synovial fluid. J Clin Microbiol 39: 4468-4471


Refbacks

  • There are currently no refbacks.